iBet uBet web content aggregator. Adding the entire web to your favor.
iBet uBet web content aggregator. Adding the entire web to your favor.



Link to original content: https://pubmed.ncbi.nlm.nih.gov/12956765
European Concerted Action on Anticoagulation. Correction of displayed international normalized ratio on two point-of-care test whole-blood prothrombin time monitors (CoaguChek Mini and TAS PT-NC) by independent international sensitivity index calibration - PubMed Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Multicenter Study
. 2003 Sep;122(6):944-9.
doi: 10.1046/j.1365-2141.2003.04521.x.

European Concerted Action on Anticoagulation. Correction of displayed international normalized ratio on two point-of-care test whole-blood prothrombin time monitors (CoaguChek Mini and TAS PT-NC) by independent international sensitivity index calibration

Affiliations
Free article
Multicenter Study

European Concerted Action on Anticoagulation. Correction of displayed international normalized ratio on two point-of-care test whole-blood prothrombin time monitors (CoaguChek Mini and TAS PT-NC) by independent international sensitivity index calibration

Leon Poller et al. Br J Haematol. 2003 Sep.
Free article

Abstract

The international normalized ratio (INR) on two widely used point-of-care test (POCT) prothrombin time (PT) monitors (CoaguChek Mini and TAS PT-NC) differed considerably and also differed from the 'true' INR obtained on the same samples using a manual PT and the same species thromboplastin international reference preparation. Agreement between the displayed INR and difference from 'true' INR has been reassessed following an independent international sensitivity index (ISI) calibration of the two systems. The displayed INRs taken at seven centres were compared with 'true' INRs from the same blood donations and INRs based on the resulting ISI. The overall difference between the displayed INRs on the two monitor systems was reduced from 21.0% to 3.5%. The overall difference in mean INR of system A from the 'true' INR was reduced from 19.0% to 9.5% and of system B from 6.8% to 0.3%, but individual centre's results still showed considerable mean INR variability. Differences between overall displayed INR with the two monitor systems have been reduced by an independent multicentre calibration, and agreement with 'true' INR on the same blood samples improved. However, marked variability in mean INR at individual centres remained after ISI correction, which demonstrates the need for external quality control of individual POCT whole-blood PT monitors.